What to do if the salt spray test box is abnormal 1. The salt spray test box cannot rise to the set temperature

1, laboratory temperature controller temperature, set too low

2, the laboratory safety switch is set too low

3, heating system failure

4, electromagnetic relay failure

5, controller failure

Approach

1, set the temperature controller to the desired temperature

2, set the safety switch to the desired temperature

3,4.5 Notice Wuhan Shangchen Test Equipment Co., Ltd.

Second, the temperature of the saturated barrel of the salt spray test chamber cannot rise to the set temperature

1, saturated barrel temperature controller temperature is too low

2, heating system failure

3, electromagnetic relay failure

4, controller failure

Approach

1, set the temperature controller to the desired temperature

2. 3.4. Notice Wuhan Shangchen Test Equipment Co., Ltd. handles

Insufficient spray rate

1, spray regulator is too low

2. Blockage of glass filter in preheating tank

3, the pressure is set too low

Approach

1, increase the spray regulator

2. Clean the glass filter

3, adjust the pressure regulator to 1kg/cm2 pressure, the compressor is marked with a pressure regulator, adjust to 2kg/cm2 pressure

Fourth, salt spray test chamber can not be sprayed

1, the air compressor is not running

2, the air compressor outlet switch did not open

3, solenoid valve failure

4, pressure gauge failure or pressure is too low

5, electromagnetic contactor failure

6, nozzle blocked

Approach

1. Open the air compressor button

2, open the air compressor main switch

3,4.5 Notice Wuhan Shangchen Test Equipment Co., Ltd.

6, remove the nozzle cleaning (carefully disassemble)

V. When the water level alarm indicator light of the salt spray test chamber is on

Indicates that the water level is too low. Check if there is water entering the inlet source.

Sixth, there is a normal spray and the air compressor is not operating

The air compressor itself has the function of self-protection and it is used as usual

Seven, can not run after turning on the power

When the water level in the warming water tank is too low, the operation power will be cut off, and the water level in the heating tank will be added to normal conditions.

Eight, salt spray test chamber temperature controller shows abnormal

Temperature controller failure, sensor failure

Belt Dryer

Materials that to be dried are distributed on the conveyor belt by feeding device. Conveyor belt will go through the channel that consists of one or several heating units. Each heating unit is equipped with air heating system and circulation system. And each channel has one or several dehumidifying system. When the conveyor belt goes through it, hot air will pass though the material that on the belt from up to down or from down to up so that the material can be dried uniformly. Heating source can be steam, heat transfer oil, flue-gas or electricity. Fan is used for heat convection and transferring heat to material, and fresh air is supplied and damp air is exhausted. In the drying time cabinet, proper relative temperature and humidity can be kept.